6
Replacing Assemblies
To remove and replace the line filter assembly
Line Filter Assembly
(16900A & 16902A)
T10 Screws (2)
82
Perform previous procedures:
•
"To prepare the instrument for disassembly"
page 48
•
"To remove and replace the cover"
Disconnect the power cable from the power supply.
1
Using a Torx T10 screwdriver, remove the 2 screws
2
securing the power socket to the frame and the 1 screw
to the ground wire.
Reverse this procedure to replace.
3
T10 Screw
Ground Wire
Line Filter Assembly
(16903A)
T10 Screws (2)
16900A, 16902A, and 16903A Service Guide
on
on page 52
T10 Screw
Ground Wire