To Remove And Replace The Front Panel Assembly - Agilent Technologies 16900A Service Manual

Logic analysis system
Hide thumbs Also See for 16900A:
Table of Contents

Advertisement

6
Replacing Assemblies

To remove and replace the front panel assembly

16900
Chassis
T10
Screws (4)
Power
On|Off
Cable
64
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 48
"To remove and replace the cover"
Disconnect the following cables from the module interface
2
board:
Front panel cable (16902/03A only)
Power On|Off cable
CD- ROM cable
3
Using a Torx T10 screwdriver, remove 4 screws that
secure the front panel to the chassis.
Remove/replace the front panel assembly.
4
Module
Interface
Front
Board
Panel
Cable
CD-ROM
Cable
Alignment
Pin
Reverse this procedure to install the front panel assembly.
5
T10
16902/03A
Screws (4)
Chassis
Power
On|Off
Cable
16900A, 16902A, and 16903A Service Guide
on
on page 52
Module
Interface
Board
CD-ROM
Cable
Alignment
Pin

Advertisement