To Perform The Power-Up Tests - Agilent Technologies 16900A Service Manual

Logic analysis system
Hide thumbs Also See for 16900A:
Table of Contents

Advertisement

To perform the power-up tests

16900A, 16902A, and 16903A Service Guide
The logic analysis system automatically performs power- up
tests when you apply power to the instrument. Any errors
are reported in the boot dialogue. Serious errors will
interrupt the boot process.
The power- up tests are designed to complement the
instrument on- line Self- Tests. Tests that are performed
during powerup are not repeated in the Self- Tests.
A monitor (16900A only), keyboard and mouse must be
connected to the mainframe to observe the results of the
power- up tests.
Exit all logic analysis sessions.
1
Press the power switch for about 1 second and release,
2
wait for the the unit to power off.
3
After a few seconds, turn the power switch back on.
Observe the boot dialogue for the following:
Ensure all of the installed memory is recognized
Any error messages
Interrupt of the boot process with or without error
message
A complete transcript of the boot dialogue is in chapter 8,
"Theory of Operation."
During initialization, check for any failures.
4
If an error or an interrupt occurs, refer to
"Troubleshooting."
3
Testing Performance
Chapter
5,
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

16902a16903a

Table of Contents