Advertisement

6
Replacing Assemblies

To remove and replace optional modules or filler panels

Locked position
50
The following procedure guides you through installing,
removing, or replacing measurement modules in the logic
analysis system frame. Information on connecting specific
multi- card measurement modules can be found in the
Installation Guide that came with your frame. Measurement
modules with different model numbers may not be connected
together in multi- card (Master/Expander) modules unless
stated otherwise. See also, the service guide that came with
your measurement module.
Perform previous procedures:
1
"To prepare the instrument for disassembly"
page 48
2
Using a 1/4- inch hex socket, loosen the thumb screws on
the modules or filler panels.
Rotate the ejector levers from the locked position to the
3
open position.
Ensure all ejectors are fully open and out of the way.
4
5
Remove filler panels or slide the modules out of the logic
analysis system frame.
Thumb screw
The following figure shows a 6- card frame. The
insertion/removal order would be the same for a 3- card
frame. A single- module configuration can be installed in
any available slot.
Open position