6
Replacing Assemblies

To remove and replace the cover

T10 Screws
C A U T I O N
52
1
Perform previous procedures:
"To prepare the instrument for disassembly"
page 48
Using a Torx T15 screwdriver, remove the handle
2
assembly.
3
Using a Torx T10 screwdriver, remove 3 screws that
secure the cover to the chassis.
Slide the cover back and up to remove.
4
T-15 Screws
Reverse this procedure to replace the cover.
5
When reinstalling the handle assembly, ensure that the screws are
torqued to 2.372 Newton meters (21 inch pounds) so that they do
not work themselves loose.