Electrical Characteristics
Power Requirements Frame(16900/902A): 115/230 Vac +/- 20%, 48-66Hz, 1300W Max
Frame(16903A): 115/230 Vac +/- 20%, 48-66Hz, 900W Max
The line voltage is autodetected by the instrument.
CAT II (Line voltage in appliance and to wall outlet)
Pollution degree 2
Environmental Characteristics
For indoor use only.
Temperature
Instrument (16900A & 16902A): 0° to + 40° C (+32° to +104° F)
Instrument (16903A): 0° to + 50° C (+32° to +122° F)
Disk Media: 10° to + 40° C (+50° to +104° F)
Probes/cables: 0° to + 65° C (+32° to +149° F)
Reliability is enhanced when operating the instrument within a temperature of +20°C to +35°C
(+68°F to +95°F).
Altitude
3,000 m (10,000 ft)
Humidity
Relative humidity 8% to 80% at 40° C (104° F). Avoid sudden, extreme temperature changes which
could cause condensation on the circuit board.
Reliability is enhanced when operating the instrument within a noncondensing humidity range of
20% to 80%. Condensation within the instrument can cause poor operation or malfunction. Provide
protection against internal condensation.
Vibration
Operating: random vibration 0-500 Hz, 10 minutes per axis, 0.3 g (rms).
Non-operating: random vibration 0-500 Hz, 10 minutes per axis, 2.41g (rms); and swept sine resonant
search, 0-500 Hz, 0.75g, 5-minute resonant dwell at 4 resonances per axis.
Weight
Model
16900A
16902A
16903A
The weight of modules ordered with mainframes will add 0.9 kg (2.0 lb) per module
Storage
Store or ship the instrument in environments within the following limits:
Temperature -40°C to +75°C (-40°F to 167°F)
Humidity up to 90% at 65° C (149°F)
Altitude up to 3,000 meters (10,000 feet)
Protect the system from temperature extremes which cause condensation on the instrument.
12
The following operating characteristics are not specifications,
but are typical operating characteristics for the Agilent
16900- series logic analyzers.
Max Net
Max Shipping
16 kg (35.2 lbs)
24.6 kg (54.2 lbs)
17.2 kg (37.8 lbs)
25.8 kg (56.8 lbs)
14.5 kg (32.0 lbs)
23.2 kg (23.2 lbs)