INFINITY COCKPIT QUICK REFERENCE GUIDE
Pulmonary Artery Wedge Pressure
1.
Select the PA parameter box or select the Procedures... button
from the main menu bar.
2.
Select the Wedge tab or button.
3.
Check scale, sweep speed, and reference waveform selections
and adjust as needed.
4.
Select the Prepare wedge button.
5.
Inflate balloon and press "Start Wedge" to begin.
o When result obtained, screen freezes.
o Deflate balloon immediately after wedge obtained.
6.
Select one of the following buttons:
o Freeze/ Adjust button to alter the PWP value manually.
o Save wedge button to save the new value.
o Cancel wedge to cancel the measurement.
Cardiac Output
BEFORE INJECTING FOR A CARDIAC OUTPUT
1.
Dräger recommends pre-filled syringes or a closed injectate
system placed in an ice bath maintained between 0°C (32°F)
and 5°C (41°F). The accuracy of measurements done with the
thermodilution method increases as the temperature of the
injectate approaches 0°C (32 °F).
2.
If using an injectate at room temperature, use a 10 cc. injec-
tate volume, unless clinically contraindicated, to get the best
results.
3.
Select C.O. parameter box.
4.
Verify correct catheter size and type. The computation con-
stant is already entered.
5.
If either one is not correct,
o Select Catheter type and choose BD / Ohmeda, Edwards /
Baxter, Arrow, or Other.
THIS GUIDE IS INTENDED FOR REFERENCE ONLY. REFER TO INSTRUCTIONS FOR USE MANUAL.ADDITIONAL INFORMATION