19501 Series Fixture B195000/B195001/B195002 Quick Start Guide
5.2

Fixture Function and Installation

The functions of the B195002 front terminals are shown in the figure
below.
(1) RTN/Low
(2) Low Contact
(3) HV Contact
(4) HV
(5) Micro switch
14
:
This terminal connects to the RTN/LOW
terminal on the high voltage module of the
19501 partial discharge tester, which is a low
potential terminal. This terminal is almost
equal to the ground terminal on the chassis.
:
This terminal connects to the HVCC-RTN
terminal on the high voltage module of the
19501 partial discharge tester.
:
This terminal connects to the HVCC-HV
terminal on the high voltage module of the
19501 partial discharge tester.
:
This terminal connects to the HV terminal on
the high voltage module of the 19501 partial
discharge tester. Do not touch during testing
and when the DANGER light is on.
:
This switch connects to the Interlock terminal
on the fixture rear side. When the acrylic