SECTION 6
OPTIONS
6-6-3

SEND DATA CONFIGURATION SECTION

This section contains the par ameter s for setting the Tr ansmitter Por tion of the BER
Meter .
Functions and par ameter s for the Tr ansmitter por tion of the BER Meter ar e as follows:
31. Send Data O utput
Displays selected O utput for BER Meter .
Selections include RF G ENERATO R and
AUD/DATA GEN OUT. Selecting RF
G ENERATO R means an RF Car r ier is modulated
with the Data Signal befor e tr ansmission.
Selecting AUD/DATA G EN O UT means Data
Signal is tr ansmitted dir ectly thr ough the
AUDIO /DATA G EN Connector .
32. RF Field
Displays RF G ener ator Fr equency. Selections r ange fr om 0.0000 to
1000.0000 MHz.
33. Level
Displays RF G ener ator output Level. Selections r ange fr om - 130 to - 30 dBm with
T/R Connector selected as O utput ( 34) . Selections r ange fr om - 130 to - 13 dBm
with AUX RF O UT Connector selected as O utput ( 34) .
34. O utput
Displays O utput Connector .
and T/R ( T/R Connector ) .
35. Modulation Type
Displays Modulation Type. Selections include O FF, AM, FM and PM.
36. Modulation/Deviation Level
Displays Modulation Level. Selections r ange fr om 0.0% to 100% for AM
Modulation Type ( 35) , 0.00 to 100 kHz for FM Modulation Type ( 35) and 0.00 to
10 Rad for PM Modulation Type.
37. O utput Level
Displays O utput Level of Data Signal.
Selections r ange fr om 0.0 to 2.5 Vp in x1
( times 1) Mode or 0 to 25 mVp in /10 ( divide
by 10) Mode.
Selections include AUX ( AUX RF O UT Connector )
6- 8
Send Data: RF GENERATOR
RF:
10.0000
Level:
-30.0
Output:
T/R
Type:
FM
Deviation:
5.0kHz
00607097
Send Data: AUD/DATA GEN OUT
Level:
7.96 Vp
00607098
31
32
33
34
35
36