Fluke 6080A Service Manual page 18

Synthesized signal generator
Table of Contents

Advertisement

INTRODUCTION AND SPECIFICATIONS
FREQUENCY (10-DIGIT DISPLAY)
RANGE
RESOLUTION
ACCURACY
REFERENCE (Internal)
REFERENCE (External)
AMPLITUDE (3 1/2-DIGIT DISPLAY)
RANGE
RESOLUTION
ACCURACY
SOURCE VSWR
FLATNESS
1-4
Unless otherwise noted, the following performance is guaranteed over the specified
environmental and AC power line conditions two hours after turn-on.
BAND
.50-15 MHz
BAND
15-32 MHz
BAND
32-64 MHz
BAND
64-128 MHz
BAND
128-256 MHz
BAND
256-512 MHz
BAND
512-1024 MHz
Table 1-3. 6080A/AN Specifications
NOTE
0.50 to 1024 MHz in 7 bands:
0.50 to 14.999999 MHz,
15 to 31.999999 MHz,
32 to 63.999999 MHz,
64 to 127.999999 MHz,
128 to 255.999999 MHz,
256 to 511.999999 MHz,
512 to 1024 MHz.
1 Hz
Same as reference (See REFERENCE).
The unit operates on an internal 10 MHz
Temperature Compensated Crystal Oscillator
(TCXO). The frequency variation will be < 10 ppm
peak to peak over the temperature range of 0 to
+50°C.
Internal reference signal (10 MHz) available at rear
panel REF OUT connector, level > 0 dBm, terminated
into 50 ohms.
Frequency stability after 2 hour warmup is < ± 0.05
ppm/hour at + 25°C ± 5°C.
Accepts 5 or 10 MHz signal. Level required is 0.5 to
2.0V RMS into 50 ohms termination.
+13 to -137 dBm
0.1 dB (<1% or 1 nV in Volts).
Annunciators for dB, dBm, V, mV, μV, dB mV,
dB μV, dBf, and EMF.
± 1.5 dB from +13 to -117 dBm
± 3 dB from -117 to -137 dBm
< 1.5:1 for levels below -10 dBm, < 2.5:1 elsewhere.
±1.0dB@+10dBm.