OPTO-EDU A10.1018 Instruction Manual

Professional dark field microscope

Advertisement

Quick Links

A 1 0 . 1 0 1 8
Professional Dark Field Microscope
Instruction Manual
Please read this manual carefully before use microscope!

Advertisement

Table of Contents
loading

Summary of Contents for OPTO-EDU A10.1018

  • Page 1 A 1 0 . 1 0 1 8 Professional Dark Field Microscope Instruction Manual Please read this manual carefully before use microscope!
  • Page 2: Use Notices

    Use Notices A10.1018 I、Safety Note 1. Carefully open the box, avoid the accessories, dropping to ground and being damaged. 2. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environment. Make sure the stage is smooth, horizontal and firm enough.
  • Page 3: Name Of Components

    1. Name of Components A10.1018 Eyepiece Video Port (Optional) Trinocular Viewing Head Nosepiece Objective Mechanical Stage Lamp House Microscope Body Installation A10.1018 Installation Diagram The following figure shows the installation sequence of the components. The numbers in the figure show the installation steps. Before installing, be sure every components is clean, do not touch glass surface.
  • Page 4 Video Port (optional) ⑥ ⑤ 10×Wide Field Eyepiece Trinocular Viewing Unit ② ④ Infinite Plan Objective ① ③ Microscope Body Mechanical Stage Support Device Lamp House...
  • Page 5: Installation Steps

    2. Installation Steps A10.1018 Guide Board Locking Block and Bolt 2.1 Installing the Mechanical Stage Support Device  Before installing the device, be sure to adjust the coarse focus knob. Make the guide board (see figure 1) down to the lowest position, so you can install the mechanical stage support device...
  • Page 6 A10.1018  Hold on the mechanical stage support device (figure 2), place it from the top of the guide board (figure 1), let the device (figure2) falling free until it reach the limit position. Use the hexagon wrench screw down the locking block; make the stage support device (figure1) and the guide board fixed together.
  • Page 7 A10.1018 2-2-4 Installing the Lamp House  Keep the bolt on the lamp house (figure 6) in line with the jack on the back of the microscope (like the show of figure 7), then pushing the lamp holder into the illumination kits gently until they are against each other (figure 8).
  • Page 8 A10.1018 2-2-6 Installing the Eyepiece Eyepiece Insert the eyepiece (figure 11) into the eyepiece tube until they are against each other. The result is showing in the figure 2-2-7Installing Video Port (optional) Bolt Insert the video port (figure 12) into the trinocular unit (figure 13), then screw down the bolt to fix it.
  • Page 9 3. Adjustment A10.1018 Video Port (optional) Diopter Ring Clocking Set Left Fine Focus Knob Left Coarse Focus Knob Condenser Focus Knob...
  • Page 10 A10.1018 Interpupillary Distance Indicator Video Port (optional) Light Path Selector Lever Swing out Condenser (with Aperture Diaphragm) Right Coarse Focus Knob Field Diaphragm Portrait Adjustment Knob Lateral Adjustment Knob Tension Adjustment Brightness Adjustment Main Switch Right Fine Focus Knob Collar Knob Note : the video port is optional.
  • Page 11: Operation

    4. Operation A10.1018 4-1 Turning on the Lamp (Figure 15) Connect the power, turn on the main switch (figure 15) to “-”(on). 4-2 Adjust Brightness (Figure 16) Turning the brightness adjustment knob Main switch clockwise, the voltage raise, and the brightness strengthen;...
  • Page 12 A10.1018 4-4 Placing Specimen (Figure 18) Place the slide on the mechanical stage. Use the stage clips to clamp the slide gently. Turn the portrait and lateral adjustment knob of the mechanical ruler, move the specimen onto the required position.
  • Page 13 A10.1018 4-7 Focus (Figure21, Figure22) 1、 When not using the video set Push in the light path selector lever (figure 25) completely, then observe with both eyes. Use the 10×objective focus, to avoid the objective touch with specimen, should raise...
  • Page 14 A10.1018 Note: 1. Using the Swing out Condenser When using the low magnification objective, turn out the condenser, and let it away from the light path. While using the high magnification objective, turn it into the light path. 2. Adjusting the Aperture Diaphragm The aperture diaphragm is designed for the adjustment of the numerical aperture , not for the brightness.
  • Page 15 5.Dark Field Function Manual A10.1018 1. Select 10x eyepiece and 10x objective to observe the specimen. 2. Place a specimen on the stage (Blood slide) 3. Replace with dark field condenser, Position of condenser is maximized. With oil immersion condenser, place a drop of oil on the front lens of objective. Slowly move the objective into place and carefully observe the oil contact area between the objective front lens and the microscope slide.
  • Page 16: Technical Specifications

    6.Technical Specifications A10.1018 Objective Specifications Numerical Working Conjugate Magnification Thickness of Magnification Value Aperture Distance Distance Sign Cover Slip (mm) (mm) Diaphragm(N.A) (Color loop) ∞ 0.10 25.42 0.17 ∞ 0.25 0.17 Yellow ∞ 0.65 0.75 0.17 Baby Blue ∞ 100X 1.25...
  • Page 17 Adjust the filament position ,let the light distributing of the field of 5、 The image The illumination light incline serious focus surface view become symmetrical and bright incline(one side is clear and the Put the specimen on the right The specimen don’t correctly place other side is position faint)