Danfoss SonoMeter 40 Calibration Instruction page 4

Ultrasonic energy meter
Hide thumbs Also See for SonoMeter 40:

Advertisement

User Guide
1.2. Activation of the test
mode by short-circuiting
the contacts (continuous)
4 | BC389148615526en-010101
SonoMeter 40
Display readings in the test mode
ID
Parameter
4.1
High- resolution energy
4.2
High-resolution integrated volume
Supply heat carrier temperature
4.3
value
Return heat carrier temperature
4.4
value
4.5
Temperature difference
4.6
High-resolution flow rate
To activate energy pulses output
(when volume pulse output is active)
4.7
To activate volume pulse output
(when energy pulse output is active)
4.8
To deactivate the test mode
In this mode:
- volume pulses are generated via the optical interface of the meter. The button can be used to
toggle the energy pulse output by selecting the menu item "tEST on Wh";
- When the meter is supplied with a connected pulse input/output cable , the energy pulses are
generated in the 1st pulse output and the volume pulses in the 2nd pulse output;
- it is possible to simulate volume pulses for determination the energy measurement errors;
- it is possible to change the parameters of the meter configuration.
NOTE: when the test mode is activated by short-circuiting the contacts "SERVICE", the volume
and energy accumulated in the test mode are not added to the meter's operating mode
volume and energy readings.
Value
Notes
Updated every second.
TEST
000000.00
Wh
Indicated as "PULSE", if the energy
TEST
test pulse output is activated.
PULSE
m
Updated every second.
3
TEST
00.000000
Indicated as "PULSE m³", if the
m
3
TEST
volume test pulse output is
PULSE
activated.
1
TEST
-
0.0 °C
2
TEST
-
0.0 °C
1-2
TEST
-
0.00 °C
m
3
TEST
0.000
-
INF
Activated by pressing and holding
TEST
tESt on
Wh
the button
3
Activated by pressing and holding
TEST m
tESt on
the button
Deactivated by pressing and holding
TEST
tESt OFF
the button
© Danfoss | Energy Meters | 2021.09