1.1 Activation of the test mode with the button .......................3 1.2 Activation of the test mode by short-circuiting the contacts ................. 4 1.3 Activation of test mode with software “SonoMeter 40 Configurator”..............5 2. Determination of measurement errors of the meter ..............6 2.1 Volume measuring errors determination test ......................6...
User Guide SonoMeter 40 General information This instruction is designed for the ultrasonic energy meter SonoMeter 40 for flow and energy verification and calibration. 1. Activation of the test The test mode can be activated in one of the following ways: mode 1.1.
Page 5
User Guide SonoMeter 40 1.3. Activation of test The test mode can be activated via the optical interface using the software “SonoMeter 40 mode with software Configurator” and optical scan head in accordance with EN 62056-21 standard. In this case, optionally “SonoMeter 40...
Page 7
Meter calibration/adjustment allows to adjust the meter measurement of volume single point adjustment mode characteristics. It can be done by using the software “SonoMeter 40 Configurator” and optical scan head in accordance with EN 62056-21 standard. 1) The calibration/adjustment mode can be activated by removing protecting lid (2) of ADJ and shortcutting the pins.