Using the Delay mode of the pulse generator channel 1, position the pulses
2
according to the setup/hold combination selected, +0.0 ps or -100 ps.
a On the Oscilloscope, select [Define meas] Define ∆ Time - Stop edge: rising.
b In the oscilloscope timebase menu, select Position. Using the oscilloscope knob,
position the falling edge of the data waveform so that it is centered on the display.
c On the oscilloscope, select [Shift] ∆ Time. Select Start src: channel 1, then select
[Enter] to display the setup time (∆ Time(1)-(2)).
d Adjust the pulse generator channel 2 Delay until the pulses are aligned according the
the setup time of the setup/hold combination selected, +0.0 ps or -100 ps.
Select the clock to be tested.
3
a Touch the clock field to be tested and then select the clock as indicated in the table.
The first time through this test, select the top multiple-edge clock.
Clocks
J
K
L
M
b Touch Done to exit the Master Clock menu.
To test the single-clock, multiple-edge, state acquisition
Testing Performance
3–49