Replacing Assemblies
To replace the module
With the probe cables facing away from the instrument, slide the module
3
approximately halfway into the mainframe.
Slide the complete module into the mainframe, but not completely in.
4
Each card in the instrument is firmly seated and tightened one at a time in step 6.
Position all cards and filler panels so that the endplates overlap.
5
Seat the cards and tighten the thumbscrews.
6
Starting with the bottom card, firmly seat the cards into the backplane connector of the
mainframe. Keep applying pressure to the center of the card endplate while tightening the
thumbscrews finger-tight. Repeat this for all cards and filler panels starting at the bottom and
moving to the top.
Correct air circulation keeps the instrument from overheating. For correct air circulation,
C A U T I O N
filler panels must be installed in all unused card slots. Keep any extra filler panels for future
use.