Integration with Arial
Integration with Arial is supported in WanderGuard BLUE v1.0 by connecting
Inovonics transmitter(s) to the Controller's relays or outputs.
Integration via Inovonics 54350:
Can be connected to dry contact (Relay 1 or Relay 2 – for example to NO and
COM)
Transmits an Alarm when 12 VDC is sent
Sends a clear message when the 12 VDC is removed
54350 can also be used for output activation:
Transmits an Alarm when the 100 mA output is activated
Sends a clear message when the 100 mA output is deactivated
Integration via Inovonics 14390:
Can be connected to 12 V relay and to 100 mA output (the 12 V can come
from the Controller)
Transmits an Alarm when the 100 mA output is activated
Sends a clear message when the 100 mA output is deactivated