Commissioning; Start Up; Operating The Device - Siemens SITRANS FUS060 Operating Instructions Manual

Ultrasonic flowmeters with profibus pa
Hide thumbs Also See for SITRANS FUS060:
Table of Contents

Advertisement

Commissioning

Dangerous high voltage
Certain parts inside the device carry dangerous high voltage. The housing must be closed and
grounded before switching the device on.
Operation with high pressure and corrosive media
The sensor connected to this device can be operated with high pressure and corrosive media.
Therefore improper handling of the device can lead to serious injuries and/or considerable
material damage.
6.1

Start up

Upon power-on the device runs a self-test which lasts about 30 seconds. Immediately after the
self-test, the device is ready for operation. Be aware that a number of factory settings (for
example max. volume flow, angle of paths, distance between transducers, and displacement of
transducer from center line of pipe) are dimension dependent. The settings are stored from
factory in the transmitter for SONO 3100 and SONO 3300. For SONOKIT and spare part
transmitters, enter values manually; see Keying in sensor data (Page 72). If there is still gas/air
in the pipe (metering tube) or in the pipeline after assembly, a flashing "F" or "D" can appear at
the top right of the first line in the display, see Diagnostics (menu 2) (Page 53) and Table 9-1
Error symptoms (Page 75). The failure signal is output at the output.
6.2

Operating the device

The device can be operated in the following ways:
• Local display (LUI)
• PROFIBUS PA
• SIMATIC PDM (PC/laptop)
SITRANS FUS060 with PROFIBUS PA
Operating Instructions, 09/2021, A5E02225054-AD