Advertisement

Maintenance and service
October 2022
5. Unplug the wiring harness plugs from the connectors J1, J2, and J3 located on the
6. Tag and disconnect the wiring to connector J4.
7. Remove the DR board from the Rosemount Xi enclosure.
8. Partially slide the new DR board into the mating slots in the Rosemount Xi
9. Refer to the wiring diagram in
10. Connect the wiring harness plugs to connectors J1, J2, J3, and J8.
11. Slide the DR board completely into the Xi enclosure.
12. Swing the Rosemount Xi cover up and tighten the four mounting screws.
74
top, right-hand side of the board. Refer to
Figure 6-9: DR board wiring, left-hand side
enclosure. Make sure the board is correctly aligned in the slots.
Figure
6-9.
Figure
6-8. Reconnect the wiring to connector J4.
Reference Manual
00809-0100-4889
Rosemount Xi