Programmable DC Power Supply 62000L Series User's Manual
If you calibrate the power supply using the remote interface, you can also
abort a calibration by placing a command via the remote interface or
pressing the Local button on the front panel.

3.12.5 Calibration String

You can leave a string before/after the calibration is completed. Usually used this string to
record the last calibration date or the next calibration due date. The calibration string contains
max 35 characters.
To read the calibration message, you can go through it via the front panel operation and the
remote control. To save the calibration message, you have to operate via the remote control
only.
Front Panel Operation:
To get the calibration information on the 62000L series, please follow the procedures below.
1. Press the
button once. Rotate the knob to the CAL.
Menu
2. Press the
button once to confirm. Rotate the knob to STRING.
Menu
3. Press the
button once to confirm. The calibration information in a scrolling text will
Menu
show on the display.
4. Use the
or
5. Press any other button to exit.
Remote Operation:
 The power supply has to be unsecured before setting the calibration string.
 The calibration string is limited up to 35 characters.
CAL:STR <quoted string>
Save the calibration message. For example, use the command CAL:STR "CAL 14-09-13" to
store the last calibration date or the next calibration due date.

3.13 Remote Control Settings

To operate the power supply through remote interface you must install communication card is
necessary. For more information about programming the power supply through the remote
control, refer to the Chapter

3.13.1 Remote Interface Selection

The USB (Model: PWR-opt01) and GPIB + USB (Model: PWR-opt02) interfaces are optional
accessories for the power supply. When using these interfaces, you are allowed to leave both
cords on the USB's and GPIB's interfaces between your power supply and the computer at
the same time. You don't need to disconnect either connection.
3-30
buttons to move forwards/backwards, or stop the running information
"Remote Control
Reference".