Programmable DC Power Supply 62000L Series User's Manual
OUTPut:SEQuence:STEP:RAMP {<step>},{<time in ms>|MIN|MAX}
Set ramp time of a step. Step is S0~S7. The ramp time allows you to set from 0 to
3599999ms.
OUTPut:SEQuence:STEP:RAMP? {<step>}
Query ramp time of a step. Step is S0~S7. Return ramp time in ms.
OUTPut:SEQuence:STEP:DWELl {<step>},{<time in ms>|MIN|MAX}
Set width of a step. Step is S0~S7. Maximum dwell is 86399999ms.
OUTPut:SEQuence:STEP:DWELl? {<step>}
Query width of a step. Step is S0~S7. Return dwell in ms.
OUTPut:SEQuence:STEP? {<step>}
Query a step setting. Step is S0~S7.

3.11 System Operations

This section provides information about system such as error message querying, beep
settings, display control, OCP delay setting, Style and power supply identity querying. Most of
the items in System Operation provide information for servicing procedures.

3.11.1 Error States

Once an error occurs, the ERR indicator is lit. Errors can be generated by invalid operation or
hardware failure. A record with up to 32 errors is saved in the power supply's error queue. For
more information, please refer to the following instructions and the chapter
 Errors are retrieved in a first-in-first-out (FIFO) sequence. The first returned error is the
first saved error. When all of errors are read from the queue, there will be no ERROR
symbol shown on the display.
 When more than 32 errors have occurred when operating the power supply over the
remote control, the last error saved in the queue (the most recent error) will be replaced
with -350,Too many errors. After removing the errors from the queue, you can save the
next error. If there is no error when you read the error queue, the power supply shows
"NO ERROR" on the display, or responds with +0,No errors over the remote control.
 You can clear the error queue when turning off the power supply or after commanding a
*CLS (Clear Status). Please note that the *RST (Reset) doesn't clear the error queue.
Front Panel Operation:
To check and remove error message(s), please follow the procedures below.
1. Press the
Menu
2. Press the
Menu
3. Press the
Menu
and SETTINGS CONFLICT at the bottom line in scrolling text. If there are still other error
message(s), repeatedly pressing the
the queue. After the last press, NO ERROR will appear on the display.
Remote Operation:
3-22
button once. Rotate the knob to the SYSTEM.
button once to confirm. A blinking ERROR will show on the display.
button once to check the error message, such as -221 at the upper line,
button will read and clear the errors stored in
Menu
"Error
Messages."