Programmable DC Power Supply 62000L Series User's Manual
Unit 1, set the master condition to the output off state.
Unit 2, set the slave condition to the output off state.
Unit 3, set the slave condition to the output off state.
After enabling the output on the master unit, the display on the master unit shows a voltage
value and a current value. On the display of the slave units, only a voltage value is shown.
Unit 1, set the master condition to the output on state.
Unit 2, set the slave condition to the output on state.
Unit 3, set the slave condition to the output on state.

3.8.3 Other Settings for Multiple Connections

● DELAY: Before setting the DELAY, you have to disable the output. This Delay allows you
to set a delay time for the output/control timing when the 62000L is/are set to the master or
normal mode in a sub-system. When the 62000L is/are set to the master or normal mode in a
sub-system. When the main system's output is enabled, the 62000L sub-system's set to the
master or normal mode, will be also activated according to the delay time (at most 60
seconds). In the meantime, the
time is up and the display returns to the meter mode. Then the slave unit(s) in the sub-system
can be controlled by the sub-system's master unit(s), and output the similar power as the
master. This sub-system application can be branched again and again to meet large system
requirements. The procedures for setting the delay time are as follows:
1. Press the
button.
Menu
2. Rotate the Knob to show MA/SL on the display.
3. Press the
button to confirm.
Menu
4. Rotate the Knob to show SETUP on the display.
5. Press the
button to confirm.
Menu
6. Rotate the Knob to show DELAY on the display.
7. Press the
button to confirm.
Menu
8. Rotate the Knob to show T_ON on the display.
9. Press the
button to set a desired delay time.
Menu
10. Press the
Menu
To quit the configuration, press any other key to exit.
● P-MA (Master Settings in Parallel): For parallel systems, the master settings can also be
set the same on the other 62000L for sub-system applications. When the main system's
output is enabled, the 62000L sub-system, set to master mode, will be also activated in a few
seconds according to the delay time setting. The procedures for setting the P-MA mode are
described in the section 3.7.2.
● NORMAL: Before enabling the NORMAL mode, you have to disable the output. Then
select the normal mode through the following procedures. Enabling this mode requires that
you remove the output connections to the other 62000L. The unit(s) under the normal mode
will only have a single output. The control connection from the master unit can exist and wait
for the order to output. When enabling the master's output, the unit(s) under the normal mode,
after the defined delay time, will go into the meter mode, and limit their output to the defined
voltages and currents according to the limit settings. To check the limited value, press the
button once. When the slave function is activated under the NORMAL mode, the
Limit
maximum output value on each slave unit follows the present adjustment.
3-16
OFF
annunciator at the bottom right will blink until the delay
button to finish the settings.