You can set a string (up to 49 characters) to show on the display. Limited by the display
digits, the string is scrolling if it is longer than 6 characters. Only alphanumeric characters
and '-' can be displayed and all letters will be change to uppercase letters. Characters
that can not be shown will be replaced with a space.
The display is enabled again and the displaying string is cleared while the power supply
is back to local operation.
Remote Operation:
DISPlay {OFF|ON}
Disable/enable the display.
DISPlay:TEXT <quoted string>
Display a string in the screen. The string must be enclosed in quoted. For example,
DISPlay:TEXT "HELLO WORLD" will show "HELLO WORLD" on the screen
DISPlay:TEXT:CLEar
Clear the displayed message.

3.11.4 OCP Delay (OCPDLY) Settings

If the OCP level is set to a value less than the CC limit, the inrush current upon output turning
on can easily cause an OCP trip. To avoid the current output being blocked by frequent OCP
trips an OCP delay is necessary. The OCP delay is a time interval that OCP function is
paused upon the output is turn on. The default for the OCP delay is 0.15s. This can be
adjusted according to your application. To change the delay time, please follow the
procedures below.
Front Panel Operation:
1. Press the
button once. Rotate the knob to the SYSTEM.
Menu
2. Press the
button once to confirm. Rotate the knob to the OCPDLY.
Menu
3. Press the
button to enter OCPDLY setting.
Menu
4. Rotate the knob and
5. Press the
button once to finish the settings.
Menu
6. To quit the configuration, press any other key to exit.
Remote Operation:
SOURce:CURRent:PROTection:DELay {<ms>|MIN|MAX}
The OCP delay time can be set to a value from 0 to 9.999s.
SOURce:CURRent:PROTection:DELay?
Query the present delay time of the OCP.

3.11.5 Vrange

Through the front panel, the Vrange is allowed users to set a maximum and a minimum
values for voltage. When the values are set, the available working voltage within the set range
just can be adjusted.
Front Panel Operation:
1. Press the
button once. Rotate the knob to the SYSTEM.
Menu
3-24
or
to leave a value, such as 0.2s.