Measurement Procedure; Flow Of Preparation And Measurement - Konica Minolta CM-3600A Instruction Manual

Table of Contents

Advertisement

Measurement Procedure

○ This manual explains how to prepare the CM-3600A and how to set a specimen.
○ The CM-3600A is controlled by a PC to perform measurements.
○ For a description of measuring method using SpectraMagic™ NX
(optional), refer to the SpectraMagic™ NX instruction manual.

Flow of Preparation and Measurement

Connecting the PC
Connecting the AC adapter
Starting the PC
(starting Windows)
Starting the software
Turning the power ON
Attaching a target mask
Performing zero calibration
Performing white (100%)
calibration
Positioning a specimen
Performing measurement
Turning the power OFF
Connect the CM-3600A to the PC with the USB cable.
(P. 13)
Connect the CM-3600A to the AC outlet
with the AC adapter. (P. 14)
Turn on the PC to be used to control
the CM-3600A.
Start the software and set it for control of the CM-3600A.
Turn the power ON. (P. 15)
<Reflectance measurement>
Attach the target mask
to be used. (P. 16)
Position the zero calibration
box and perform zero
calibration. (P. 17)
Position the white calibration
plate and perform white
calibration. (P. 18)
Position the specimen on / in the CM-3600A. (P. 21)
Perform measurements.
When measurements are complete, turn the power OFF. (P. 15)
Exit out of the software and turn off the PC.
12
<Transmittance measurement>
Attach the LAV (Ø25.4 mm)
target mask and set the white
calibration plate in position. (P. 16)
Position the Transmittance
Zero Calibration Plate and
perform zero calibration. (P. 19)
Perform 100% transmittance
with nothing or a cell
containing distilled water in the
transmittance chamber. (P. 20)