SALUDA MEDICAL Evoke SCS User Manual page 40

Hide thumbs Also See for Evoke SCS:
Table of Contents

Advertisement

15.1.2 Evoke eCLS (includes Case and Lead Adapters)
Table 15-2: Evoke eCLS.
Materials
Dimensions
Weight
Electrodes
Functional
specifications
Radio frequency
communication
*Channels are automatically selected when the communication session begins.
Battery
Charging
Ingress Protection
eCLS in case
eCLS
IEC 60601-1 /
EN 60601-1
Classification
Storage &
Transportation
Conditions
Operating
Conditions
CLIN-UMAN-005176
Evoke® SCS System User Manual
eCLS body and case
Case seal
100 mm x 85 mm x 20 mm (3.9 in x 3.4 in x 0.8 in)
96 g (3.4 oz.)
24
Port 1: electrodes 1-12, Port 2: electrodes 13-24
All other functional specifications are the same as the CLS (See
Section 15.1.1)
MICS band
8 channels*
Modulation type
Range
Effective Radiated
Power (ERP)
800 mAh 3.6 V Li-ion rechargeable battery
Li-Ion battery charger
IP22 Rating for protection against access of solid objects greater than
or equal to 12.5 mm, and for vertically dripping water when the
device is tilted 15 degrees.
IP30 Rating for protection against solid objects greater than or equal
to 2.5mm, and no protection against water.
Type BF Applied Part
Internally Powered Medical Electrical Equipment
Continuous Operation
Temperature:
Humidity:
Pressure:
Temperature:
Humidity:
Pressure:
Revision: 4.01
402.45-405.55 MHz
Centre frequency (MHz): 402.45, 402.75,
403.05, 403.35, 403.65, 403.95, 404.25, 404.55
Frequency Shift Keying (FSK)
1.0 m (3.3 ft.)
25 µW (-16 dBm) maximum
Min: -10 °C (14 °F)
Min: 0% RH
70 kPa (0.69 atm)
Min: 5 °C (41 °F)
Min: 15% RH
70 kPa (0.69 atm)
ABS Plastic
TPE
Max: 55 °C (131 °F)
Max: 90% RH
Max: 150kPa (1.48 atm)
Max: 40 °C (104 °F)
Max: 90% RH
Max: 106 kPa (1.05 atm)
Page 40 of 53