Charger Audio Indicators; Recharging Your Evoke Charger - SALUDA MEDICAL Evoke SCS User Manual

Hide thumbs Also See for Evoke SCS:
Table of Contents

Advertisement

11. The Charger will beep once if the Charger coil moves out of charging range – realign the
Charger coil so the 'Charging Link' indicator shows three or four bars.
12. When charging is complete, the Charger will emit a long beep (one second) and display
four bars on the 'CLS Battery Level' indicator for 10 seconds.
Caution: Turn therapy off before you start charging. You may experience increases
or decreases in therapy strength if you start charging while therapy is on. You may turn
therapy on once charging has started, however, closed-loop stimulation is disabled
during charging and you may experience increases or decreases in therapy strength. In
some cases, the therapy may stop during charging.
Caution: When the orange bar on the 'Stimulator Battery Level' indicator on the
EPC is displayed, you should charge the CLS or your stimulator will turn off and you will
not receive therapy. If the stimulator turns off, you should charge the CLS as soon as
possible.
Note: You must unlock the EPC before the 'Stimulator Battery Level' indicator
illuminates.

8.3 Charger audio indicators

In addition to its visual indicators, the Charger will beep to indicate various functions:
• A short beep will occur every time you press a button.
• A long (one second) beep, accompanied by all the bars glowing on the 'CLS Battery
Level' indicator for 10 seconds, indicates that the charge session is complete.
• A long (one second) beep where the battery level indicator does not illuminate,
indicates that the charge session was aborted by the user.
• Three beeps indicates that the charge session has been aborted, and will be
accompanied by a visual indicator of the reason why:
o One bar glowing on the 'Charger Battery Level' indicator shows that Charger
battery is too depleted to charge your CLS and requires recharging before use
(refer to Section 8.4).
o One orange bar glowing on the 'Charger Link' indicator for 10 seconds indicates a
poor charge link.
• Three beeps accompanied by the 'Contact Clinician' indicator flashing on and off six
times indicates an abnormal condition and that you should see your clinician to resolve
the issue.

8.4 Recharging your Evoke Charger

Follow the steps below to charge your Evoke Charger:
1. If charging, remove the Charger coil from over the stimulator.
CLIN-UMAN-005176
Evoke® SCS System User Manual
Revision: 4.01
Page 31 of 53

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents