instructions below. Do not pull on the leads or extensions or apply tension between the
eCLS and the leads.
5.5.4.1 Disconnecting the eCLS
If you require the eCLS to be disconnected from the lead adapters, a care-giver should:
1. Turn stimulation off (see Section 5.5.3).
2. Open the eCLS case by separating the tabs on the eCLS case.
3. Remove the eCLS from the case.
4. Remove the lead adapters from the eCLS (see Figure 5.2), noting the port that each lead
adapter was removed from.
5. Secure the lead adapters to your back using adhesive tape to ensure they do not pull on
the leads.
If the lead adapters have been removed, you should return to the clinic to have the
components reconnected.
Caution: Do not remove the leads from the lead adapters. Do not pull on the leads
or extensions or apply tension between the eCLS and the leads while the lead adapters
are being disconnected from the eCLS.
CLIN-UMAN-005176
Evoke® SCS System User Manual
Revision: 4.01
Page 17 of 53