SALUDA MEDICAL Evoke SCS User Manual page 39

Hide thumbs Also See for Evoke SCS:
Table of Contents

Advertisement

Radio frequency
communication
*Channels are automatically selected when the communication session begins.
Battery
*End of Battery Life is defined by Saluda Medical as the point at which the device can no
longer maintain enough charge to provide 24 hours of therapy. At higher or lower settings,
this defined end of life could be shorter or longer respectively.
Charging
Recording
amplifier gain
Data recording
Radio opaque
identifier
Storage &
Transportation
Conditions
CLIN-UMAN-005176
Evoke® SCS System User Manual
MICS band
8 channels*
Modulation type
Range
Effective Radiated
Power (ERP)
Battery type
Battery life
Transcutaneous charging using inductive coupling with an external coil
Implant depth
Low: 250x
High: 1000x
32 MB, up to 1 year
(Stimulation usage, ECAP amplitude and current statistics)
"SME BYY"
Where "SME" is Saluda Medical, "B" is the CLS model and YY is the
two-digit year of manufacture
Temperature:
Revision: 4.01
402.45-405.55 MHz
Centre frequency (MHz): 402.45, 402.75, 403.05,
403.35, 403.65, 403.95, 404.25, 404.55
Frequency Shift Keying (FSK)
1.0 m (3.3 ft.)
25 µW (-16 dBm) maximum
200 mAh Li-Ion rechargeable battery
Greater than 10 years at moderate settings
(current = 5.0 mA, pulse width = 200 μs,
frequency = 60 Hz, impedance = 750 Ω,
24hrs/day usage)*
5 mm to 20 mm (0.2 in to 0.8 in)
Min: -10 °C (14 °F)
Max: 55 °C (131 °F)
Page 39 of 53