Calibration - Fisher Scientific Isotemp 178L 120V User Manual

Peltier low temperature incubators
Table of Contents

Advertisement

Handling and Control
Settings
Table 7-9 Reading the Error Log

Calibration

The Settings -> Calibration menu item enables the user to initiate a temperature calibration
process (see
sensors and choose whether calibration should be accomplished manually or automatically:
• The Manual option allows for entering an absolute temperature directly, as measured - for
example, using an external reference sensor.
Table 7-10 Entering the Calibration Reference Temperature Manually
7-10
Peltier Low Temp. Incubators
Press
previous one).
After reaching the entry numbered
returns to the beginning of the error log, displaying
again.
To exit from the error log and return to normal display mode
press
The
Settings
"Temperature Calibration Procedure"
Maintain the ambient conditions within the specified limits of the
incubator before launching calibration.
Varying ambient conditions may impact the result of the
calibration routine, which may lead to mis adjustment of the
controller and unreliable temperature control operation.
Prepare for temperature calibration (see
ture Calibration"
page
10-3).
Press
to select the Settings icon and press
to go to the next entry (or
twice.
icon in the menu bar will go out.
on
page
10-3) for the built-in temperature
Calibration Prerequisites
and
"Comparison Measurement Procedure"
to activate the menu bar, then use
to go back to the
the display wraps and
22
E01
"Preparing Tempera-
or
to confirm.