1 Important information
1.3 Safety information
8
1.3

Safety information

As manufacturers of dental and laboratory equipment and in the interest
of the operational safety of your system, we stress the importance of
having maintenance and repairs performed exclusively by our own
personnel or through our authorized representatives. Furthermore,
safety-critical system components must always be replaced with original
spare parts.
We suggest that you request a certificate specifying the nature and extent
of the work performed from the persons who performed it. Furthermore,
the certificate should also include any possible changes in rated
parameters or working ranges as well as the date, the name of the firm
and a signature.
Modifications to this system which may affect the safety of the operator or
third parties are prohibited by law.
For reasons of product safety, this product may be operated only with
original Sirona accessories or third-party accessories expressly approved
by Sirona. The user is responsible for any damage resulting from the use
of non-approved accessories.
If any equipment not approved by Sirona is connected, it must comply
with the applicable standards:
● EN 60 950 for information technology equipment, and
● EN 61 010-1 for laboratory equipment.
CAUTION
Damage/danger resulting from unauthorized alterations
The inFire HTC speed high-temperature furnace may only be altered
with our express prior approval. We are not liable for unauthorized
alterations of any kind.
Prior to opening the unit, connecting a measuring instrument or
replacing parts:
➢ Switch off the unit.
➢ Disconnect the power plug.
WARNING
Separate circuit
The inFire HTC speed high-temperature furnace must always be
connected to a separate circuit with at least a 16A fuse. The fuses must
be of the C type.
WARNING
Risk of burns due to hot surfaces
The inFire HTC speed high-temperature furnace may develop hot
surfaces in some areas after extended use. Use the necessary caution
around the unit, especially when the furnace is operated for extended
periods.
➢ Never touch the hot surfaces.
Dentsply Sirona
Service Manual inFire HTC speed
61 30 608 D3497
D3497.076.01.09.02 10.2018