Error: Led On Control Panel Indicates Heating, But Furnace Does Not - Dentsply Sirona inFire HTC speed Service Manual

Hide thumbs Also See for inFire HTC speed:
Table of Contents

Advertisement

8 Troubleshooting
8.2 Error without message on the display
24
8.2.3
Error: LED on control panel indicates heating, but furnace
does not heat
Cause
Heating defective.
Elimination
Check whether heating is jammed, measure resistance with ohmmeter:
R
= 0.6 Ω
o.k.
individual heating element = 0.1 Ω
Replace defective heating element (see "Replacing the heating element
[ → 29]").
Call customer service.
Cause
Controller board defective.
Elimination
Replace controller board (see "Replacing the controller board [ → 30]").
Call customer service.