Dentsply Sirona
Service Manual inFire HTC speed
61 30 608 D3497
D3497.076.01.09.02 10.2018
8
Troubleshooting
8.1
Error messages from electronic components
Error messages are displayed on the LCD screen. They are accompanied
by an acoustic signal.
8.1.1
Error message: SENSOR BROKEN
Cause: Thermocouple connections are loose.
Elimination: Check the thermocouple connections.
Cause: Thermocouple is defective.
Elimination: Replace the thermocouple.
See "Replacing the controller board [ → 30]".
8.1 Error messages from electronic components
8 Troubleshooting