Technical Data - Dentsply Sirona inFire HTC speed Service Manual

Hide thumbs Also See for inFire HTC speed:
Table of Contents

Advertisement

1 Important information
1.2 Technical data
Model designation:
Type of protection against electric shock:
Degree of protection against ingress of water:
Setup location:
Height
Temperature range:
Relative humidity
Overvoltage category:
Pollution degree:
Mode of operation
Dimensions of unit (W x H x D in mm):
Dimensions of packaged unit (W x H x D in mm):
Furnace chamber:
Maximum sintering temperature:
Approx. weight without packaging:
Approx. weight including packaging:
Maximum inlet pressure:
(applies to all devices with built-in gas management
module)
Rated line voltage
(all units except REF: 6401462 and 6416247):
Rated line frequency
(all units except REF: 6401462 and 6416247):
Rated current consumption
(all units except REF: 6401462 and 6416247):
Main unit fuse F1/F2
Fuse F3 control board emergency cooling
Fuse F4 control board emergency cooling
6
1.2

Technical data

inFire HTC speed
Protection Class I device
Ordinary device (without protection against ingress of
water)
Indoors in dry area
up to 2000 m
5 °C to 40 °C (41 °F to 104 °F)
80% up to 31 °C (87.8 °F), above this decreasing linearly
to 50% at 40 °C (104 °F), no condensation
II
2
Continuous operation
500 x 820 x 565
630 x 1300 x 730
Diameter 130 mm x 80 mm height
1,650 °C (3,002 °F)
80 kg
96 kg
8 bar
200 V AC – 240 V AC
50/60Hz
15 A
for 200 VAC − 240 VAC: T16A,
Order No. 61 32 448 (10 pcs)
for 250VAC: T200mA,
Order No. 61 32 463 (10 pcs)
for 250VAC: T2.5A,
Order No. 61 32 471 (10 pcs)
Dentsply Sirona
Service Manual inFire HTC speed
61 30 608 D3497
D3497.076.01.09.02 10.2018