8.2.9
Error: Furnace triggers circuit breaker
Cause
Unsuitable circuit breaker.
Elimination
Install a 16 A K circuit breaker.
Use circuit breaker type C from serial no. 3000.
8.2.10
Error: Furnace triggers the circuit breaker in the fuse box
Cause
Heating defective.
Elimination
Check whether heating is jammed, measure resistance with ohmmeter:
R
= 0.6 Ω
o.k.
individual heating element = 0.1 Ω
Replace defective heating element (see "Replacing the heating element
[ → 29]").
Call customer service.
Cause
Fault current circuit breaker is too sensitive.