Country-Specific Adaptation Of The Maximum Tube Current - Siemens ARCADIS Varic Start-Up Instructions

Hide thumbs Also See for ARCADIS Varic:
Table of Contents

Advertisement

36
USA
Service mask
input in 1/10 min.
Europe
Service mask
input in 1/10 min.
Make country-specific changes and click <Save>.
Click <Home>.
Perform a system shutdown.
Switch on the system and wait for the system to boot.
NOTE
If a change is made to "Block time settings", check this function.

Country-specific adaptation of the maximum tube current

NOTE
Start service program and log in as described under "Service Login".
ARCADIS Varic
"Warning time limit"
5 min.
50
5 min.
50
With the introduction of the Hibernation Mode, system configura-
tion changes are not applied if the system is switched off and then
switched on again using the power buttons on the monitor cart.
To correctly apply the modified data, perform a system shutdown.
Open the <Options> - <End Session> menu and click the "Shut-
down system" button. Immediately thereafter, press the Power off
button on the monitor cart.
The system shuts down and switches off automatically.
A maximum tube current limit is required - as far as known - only
in Denmark.
SPR2-310.815.01.07.02
12.06
Specific system settings
"Blocking time limit"
10 min. (5 min. + 5 min.)
Page 36 of 64
CS PS SP
0 min.
0
50
Siemens AG
Medical Solutions
0