General information
WARNING
The system must be switched on during measurement. Measuring devices with auto-
mated measuring sequences must therefore be set to manual measurement.
Enter the highest value as the first measured value in the system leakage current proto-
col.
This value must not exceed the permissible leakage current values according to DIN VDE
0751-1/2001-10, Table F.1, line "system leakage current for units according to remarks 1
and 3" of
2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a measure-
ment at the nominal value of the line voltage. This must also be documented.
Document the measuring procedure (differential measurement or direct measurement)
and the measuring device used (designation and serial number).
Separate the protocol sheet from these instructions and attach it to the protocols already
in the system binder or log book.
Repeat measurement
When maintenance or repair work is performed on the primary power supply circuit (e.g.
repairs to the power-on circuit or replacement of the line filter), the system leakage current
measurement must be repeated.
The same measuring conditions as in the first measurement apply.
Record and evaluate the highest value determined in the repeat measurement in the
existing system leakage current protocol.
This value must not exceed the permissible leakage current values according to DIN VDE
0751-1/2001-10, Table F.1, line "system leakage current for units according to remarks 1
and 3" of
2.5 mA.
Measure and record the current line voltage. If the measured line voltage deviates from
the nominal voltage, correct the measured value to the value corresponding to a measure-
ment at the nominal value of the line voltage. This must also be documented.
Document the measuring procedure (differential measurement or direct measurement)
Siemens AG
Medical Solutions
Electrical voltage!
Non-compliance can lead to severe injury and even death.
No housing parts of the system may be touched during
direct measurement of the system leakage current (mea-
surement setup according to
Third-person access to the system must be prevented.
SPR2-310.815.01.07.02
12.06
(Fig. 3 / p.
Page 17 of 64
CS PS SP
17
15)).