General information
Fig. 2:
Pos. 1
Pos. 2
Pos. 3
If the direct measurement of the system leakage current is used (measurement setup
according to
must not be touched.
Fig. 3:
Siemens AG
Medical Solutions
Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
= System
= Application part type B (if available)
= Measurement setup (integrated into measuring device)
(Fig. 3 / p.
15)), the system must be insulated during the measurement and
Measuring circuit for direct measurement of the system leakage current in compliance
SPR2-310.815.01.07.02
12.06
Page 15 of 64
CS PS SP
15
ARCADIS Varic