Hamamatsu C7557 Instruction Manual page 20

Mcd controller
Table of Contents

Advertisement

HAMAMATSU
Timing chart
Control external shutter
For control of an external shutter or an external lump, this system outputs a
pulse signal from the "PULSE OUT" connector. See timing chart below.
Timing chart of PULSE. OUT
Timing chart of PULSE. OUT ( SHUTTER DURATION is 0 )
PULSE. OUT
The PULSE OUT is available for dark current data aquisition. When the sh utter 
con trol bu ttton   "SHUTTER ON" pressed, the PUL SE OUT b ecomes High-level. 
When the shutter contro l buttton "S HUTTER OFF" pressed, the PULSE OUT 
becomes Low-level. 
Timing chart of PULSE. OUT
Timing chart of PULSE. OUT ( SHUTTER DURATION is not 0 )
DATA  OUT
PULSE.   OUT
When the sh utter control b uttton "SHUTTER ON" pressed, the PUL SE OUT o utputs 
the pulse signal. Set d elay time and p ulse width from the application software. Setted 
p ulse width is equ al to expsure time of the multichannel d etector.
( SHUTTER DURATION is 0 )
SHUTTER ON
( SHUTTER DURATION is not 0 )
Delay Time
Pul se Width
( E xposuretime)
- 19 -
MCD CONTROLLER C7557
Chapter1 Hardware
SHUTTER OF F
No t avairable data
A vairable data
Delay Time