Timing chart
External synchronization measurement
When measuring the spectral of a flash lamp, laser, and other pulsed light
with this system, use the external synchronization measurement mode. See
timing chart below.
Timing chart of external synchronization measurement
DATA  OUT
TRIG. IN
When the TR IG. IN input at scan time, the scan are val id.
When the TRIG. IN input at blank time, the next scan are val id.
Method
1) Connections
2) Operation
Timing chart of EXTERNAL TRIGGER MODE
1EVENT
Input the trigger signal to the TRIG. IN terminal at
TTL/High impedance using a BNC-BNC cable.
Set Trigger MODE to External and the Number of
Cycle from the application software. Next measure
the data using the data measurement function.
- 16 -
MCD CONTROLLER C7557
Chapter1 Hardware
Not avairabl e data