7. Exit the program by clicking on File \ Exit \ Yes.
8. Verify the program was saved by going to Start \
"Programs" \ "Accessories" \ "Communications" \
"HyperTerminal" \ "SPD210".
9. This completes the configuration of HyperTerminal.
10. Turn on SpeedVac™ system and connect to the virtual
COM port that is linked with the SpeedVac™ attached to
the computer.
11. Start HyperTerminal by clicking on SPD210.
12. Initiate the run and SpeedVac™ will screen print
<temperature>, <heat time>, <run time>,.

Lid Stay Mechanism

The unit is equipped with a hydraulic damper that will assist
the user in closing the cover. It will secure the lid and maintain
it open, allowing the user to operate the unit with both hands.
Figure 3. Lid stay mechanism view
WARNING: Device lid can crush your fingers.
Keep away: Do not reach between the device
and lid when opening or closing the device lid.

Other Topics

Chamber Cover. The cover must be closed before
beginning a run. If you press Manual Run or Auto Run and
the cover is open, the display shows "lid" to remind you to
close the cover. The run will start immediately after closing the
cover.
If a user tries to open the lid at the start of the run, when the
lid has already been locked, the unit will automatically abort
the run and signal the user with the "Lid" message on the
display and an error tone. Otherwise, during a run the user
cannot open the cover due to vacuum in the concentration
SpeedVac™ SPD210
chamber. The cover is locked down at all times during a run
and whenever power to the unit is interrupted.
The cover lock is an additional safety feature that reduces the
risk of injury or damage while rotor is spinning. Never bypass
the cover lock mechanism during a run.
If removal of samples from the concentrator chamber during
a power failure is required, insert cover lock emergency
release tool into the slot provided on the right side of the unit.
Press the object gently into the slot until the lock releases.
The cover can then be opened.
Figure 4. Opening chamber cover in emergency
CAUTION:
List of Solvent and Solvent Combinations:
Solvent & Solvent Combination
Water+ACN+MeOH+low concentration of 0.1%
Trifluoroacetic acid
Acetonitrile
Ammonium hydroxide
n-Butyl alcohol
Ethanol
n-Hexane
n-Heptane
Isopropyl alcohol
Isobutyl alcohol
Methanol
n-propyl alcohol
Pyridine
In the event that your choice of solvents and applications are
unique and not listed above, please contact Thermo Scientific
Customer Support for advice.