Thermo Scientific SpeedVac SPD210 Installation And Operation Manual page 16

Vacuum concentrator
Table of Contents

Advertisement

Appendix 2
1. TO INCREASE THE RATE OF SOLVENT EVAPORATION
AND SHORTEN DRYING TIME, TRY ONE OR MORE OF
THE FOLLOWING:
Use glass tubes and vials in place of plastic.
Preheat the SpeedVac™ chamber and rotor before
drying.
Use rotors with metal tube carriers instead of plastic
rotors (e.g., RH12-20, RH8-18).
These rotors conduct and transfer heat to tubes for
faster sample drying.
Prewash plastic tubes with methanol to remove
materials that inhibit or prevent evaporation of
aqueous samples.
2. TO KEEP VACUUM PUMP OIL CLEAN, TRY THE
FOLLOWING:
Empty solvent(s) collected in the Refrigerated Vapor
Trap as frequently as possible.
Change the oil in the vacuum pump when it
becomes brown in color or smells of solvents being
evaporated.
Attach a Thermo Scientific VPOF110 Vacuum Pump
Oil Filter/Recirculator to your vacuum pump when
evaporating acids (e.g., HCl, acetic, TFA) in the
SpeedVac™.
Attach a Thermo Scientific SCT120 Chemical Trap
with acid neutralization cartridge when evaporating
acids (i.e. HCl, acetic, TFA) in the SpeedVac™.