Chapter 4: Patient preparation
Connection of the Chest Electrodes
Press the suction bulb and place the chest electrode on proper site (Figure 4-3) and then release.
The location of the chest electrodes is as follows:
C1 (V1): Fourth intercostal space at the right margin of the sternum
C2 (V2): Fourth intercostal space at the left margin of the sternum
C3 (V3): Midway between V2 and V4
C4 (V4): Fifth intercostal space at the left midclavicular line
C5 (V5): Left anterior axillary line at the horizontal level of V4
C6 (V6): Left mid-axillary line at the horizontal level of V4

Detection of electrode disconnection

DENA 650 continuously monitors the connection status of the electrodes, and in the event of a
disconnection, displays the relevant messages at the location of the signals on the screen (Figure 1-2).
If any of the R, L, or F electrodes are disconnected, the messages Check R, Check L, and Check F
are displayed.
If any of the chest electrodes are disconnected, the message Check Cx is displayed (x from 1 to
6).
If the N electrode is disconnected, one or more disconnection messages may be displayed.
NOTE
Refer to Troubleshooting and
Figure 4-3 Connection of the Chest Electrodes
Error messages chapter to view device messages.