Checking The Compression Unit - Siemens MAMMOMAT Novation DR Parts Replacement

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Compression Unit
Repeat the measurement once/twice if the scales show a value exceeding the pre-
scribed tolerance +/- 0.3 kp of the nominal 20 kp. To do this, decompress by about 50 N
via the other foot-switch and then compress again.
If the (average) force shown on the scales doesn't fall into the given range, the hexa-
decimal value for force = 200 N stored in the data field has to be adjusted.
(This is only an approximation, meant as an example; please observe the actual
behavior: decreasing the default value "02D8" by four units (to: 02D4) will result in a
force decrease of approximately 2 N. A corresponding increase (to: 02DC) will result in
a force increase of approximately 2 N. )
Adjust the hexadecimal value and save it by pressing F2.
When the adjusted value is saved (under calibration ), cross-check the force as de-
scribed above. Adjust the hexadecimal value, if necessary. Perform another
cross-check.
Save the final value by pressing F2. This should exit the menu (otherwise press ESC).
Optimized compression
Select Optimized Compression in the main menu, under Configuration > Compres-
sion.
Re-enter the value "5137" in both OPCOMP data fields, i.e. for OPCOMP 18 x 24 and
for OPCOMP 24 x 30.
Save the entered value by pressing F2.
Exit the menu by pressing ESC.

Checking the compression unit

Maximal compression – lowest position of the paddle
Attach the same compression plate used for the work step: "calibrating the thickness
top mm" to the compression unit.
Activate the foot-switch to move the compression plate cautiously downward.
Check whether the bottom of the paddle has reached the surface of the carbon cover.
Lower endpoint
First detach and remove the compression plate from the unit.
Attempt further downward motion via the foot-switch. Keep the switch pressed until the
sled reaches its lowest limit position (mechanical stop).
Read and record the compression thickness data (in mm) for the lower endpoint, as
shown on the stand display.
Upper endpoint
Move the compression plate upward to its upper endpoint.
Reattach the compression plate and measure the distance from the top of the carbon-fi-
ber cover of the object table to the bottom of the paddle of the compression plate, e.g.
by means of a tape measure. (As for the calibration of "thickness top mm".)
Read and record the corresponding data shown on the stand display. Compare it to the
data just measured.
Remove the compression paddle.