To end the BRICK web interface session, close the web browser window.
syngo-based acquisition workstation (WH AWS)
Log on to the service software. Select Options-> Service> Local service. Enter the
service key (6 characters in the 2nd field) in the syngo software screen. Confirm the
screen with OK. Select Acquisition system. Select Brick configuration.
The BRICK home page opens in a new Internet Explorer window.
DROC acquisition workstation
Click the Netscape web browser icon on the acquisition workstation and enter the
following URL:
http://brick
The PXCM home page opens in the web-browser window.
Any time a user attempts to use a function assigned a moderate or high security level, a
login prompt is displayed to verify that he or she is authorized to use the function.
The user name and appropriate password have to be entered:
login: root
password: ******
Once you have logged into a particular function, you can use it throughout the same
PXCM web tool session. You do not have to log into that function again.
Establishing a connection to the BRICK command line interface
To access the command line BRICK interface, start a telnet session on the acquisition
workstation.
The procedures differ for opening the BRICK command line interface on the
syngo-based acquisition workstation (WH AWS) and the DROC acquisition workstation.
The BRICK command line interface session can be closed by pressing Ctrl and C simul-
taneously.
CAUTION
syngo-based acquisition workstation (WH AWS)
- Enter the following command in the syngo Service Software Utilities -> Escape To
OS command line:
start telnet brick
A command tool opens with the UNIX login.
- The user name and appropriate password have to be entered:
login: root
password: ******
MAMMOMAT Novation DR
Replacing basic unit electrical assemblies
It is very important to spell the command properly when typing it
in.
If an incorrect command is entered, the BRICK may suf-
fer damage which can be very difficult to remedy.
SPB7-250.841.01.04.02
05.05
Page 22 of 64