Replacing Stand Boards
D801 stand CPU board
Fig. 14: D801
•
Make a backup of the stand data. See Software System Stand and Backup/Restore.
•
Turn the system OFF.
•
Remove the right stand cover.
•
Move prom J9 and J65 from the old to the new Board
•
Install the new board.
•
Install the right stand cover.
•
Turn the system ON.
•
Restore the stand data.
Siemens AG
Medical Solutions
SPB7-250.841.01.04.02
05.05
(1/Fig. 14 / p.
39).
MAMMOMAT Novation DR
Page 39 of 64
CS SD 24
39
0