D801 Stand Cpu Board - Siemens MAMMOMAT Novation DR Parts Replacement

Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Replacing Stand Boards

D801 stand CPU board

Fig. 14: D801
Make a backup of the stand data. See Software System Stand and Backup/Restore.
Turn the system OFF.
Remove the right stand cover.
Move prom J9 and J65 from the old to the new Board
Install the new board.
Install the right stand cover.
Turn the system ON.
Restore the stand data.
Siemens AG
Medical Solutions
SPB7-250.841.01.04.02
05.05
(1/Fig. 14 / p.
39).
MAMMOMAT Novation DR