Electromagnetic Emissions And Immunity - Boston Scientific LATITUDE 6288 Patient Manual

Communicator
Hide thumbs Also See for LATITUDE 6288:
Table of Contents

Advertisement

Electromagnetic Emissions and Immunity

Table 2. Guidance and manufacturer's declaration—
electromagnetic compatibility
The LATITUDE Communicator is suitable for use in a professional health care
facility or home environment.
Test
Protection of radio services and
other equipment
Protection of the public mains
network
Electrostatic discharge
Radiated RF EM field
Proximity fields from RF wireless
communications equipment
Rated power frequency magnetic
field
Electrical fast transients/ bursts
Surges line-to-line
Surges line-to-ground
88
LATITUDE
Compliance
CISPR 11
Group 1
Class B
CISPR 11 Class B
IEC 61000-3-2
IEC 61000-3-3
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, and
± 15 kV air
3 V/m from 80 MHz to
2.7 GHz
380 - 390 MHz: 27 V/m
430 - 470 MHz: 28 V/m
704 - 787 MHz: 9 V/m
800 - 960 MHz: 28 V/m
1700 - 1900 MHz: 28 V/m
2400 - 2570 MHz: 28 V/m
5100 - 5800 MHz: 9 V/m
30 A/m
± 2 kV input AC power
± 1 kV SIP/SOP
± 0.5 kV, ± 1 kV input AC
power
± 0.5 kV, ± 1 kV, ± 2 kV input AC
power
Communicator Patient Manual
Electromagnetic
environment—guidance
The LATITUDE Communicator
uses RF energy only for its
intended uses in communication
with the implanted device or
connectivity functions. Its RF
emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
The LATITUDE Communicator is
suitable for use in a professional
health care facility or home
environment.