Sentry Equipment Corp | Sample Sequencer
5
NOTES:
Registers 40001 through 40006 provide status information corresponding to SEQ1.
Registers 40007 through 40012 provide status information corresponding to SEQ2.
Registers 40013 through 40018 provide status information corresponding to SEQ3.
Registers 40019 through 40024 provide status information corresponding to SEQ4.
Registers 40001 through 40024 are read only registers.
The values located in registers 40001 and 40007 represent the following states:
0 = OFF, 1 = FLUSH, 2 = SAMPLE, 3 = ALARM, 4 = CALIBRATION, 5 = BATCH
The analyzer value is stored in consecutive registers. The first register stores the value to the
left of the decimal. The second register stores the value to the right of the decimal.
The values located in the Sample/Bypass registers represent the following states:
0 = Bypass, 1 = Sample
Registers 40065 through 40072 are write only registers. To start or stop SEQ1, SEQ2, SEQ3,
or SEQ4 a 1 should be written in the corresponding register.
64
SPD 5.5.15