The Widgets are reached by touching the wrench button in the lower right hand corner of the
Control Screen. Selecting a widget will take you to its corresponding screen. Selecting the
cancel button in the lower left corner will return you to the previous screen. Selecting the arrow
button in the lower right corner will take you to additional widgets.
5.2.5
Flush Times
Use this widget to update Flush Times. In Continuous mode, Flush Time is the amount of time
the analyzer is flushed out before the analyzer reading is considered valid. The valve for the
sample point is open during both the Sample Time and the Flush Time. In Batch mode, Flush
Time is not used.
Figure 10: Flush Times Screen
To change Flush Time for a sample, touch the time that you would like to change. The Enter Time
Screen will be shown. Enter the time desired using this screen; when finished, touch the enter
button to update the value in the Flush Times Screen, or press the escape button to return to the
Flush Times Screen without entering the value. Time values are limited to 59 minutes and 59
seconds. Once you have updated all the Flush Times, press the save button to save all times on
the screen, or touch the escape button to leave the screen without saving the updated values.