3 QUICK START GUIDE

The following guide is intended to help you get started using the Sentry Sample Sequencer
This guide does not include detailed explanations and should be used in conjunction with the
remainder of this manual to build a full understanding of the Sample Sequencer.
3.1

Electrical Connections

Connect all electrical connections necessary for the application. If the analyzer being
shared is a batch type analyzer, an end of analysis signal must be connected to the
Sample Sequencer "End of Batch" input. Without this signal, the Sample Sequencer will
never advance to the next sample.
3.2

Control Screen

With the display showing the Operation Screen (see Figure 6), touch anywhere on the
screen excluding the left and right arrow areas. If prompted, enter password. This will
take you to the Control Screen (see Figure 8).
3.3

Starting and Stopping the Sample Sequencer

From the Control Screen (see Figure 8), you can start and stop any active sequence by
touching the red indicator on the screen. Once touched, the indicator will change and the
sequencer state will update.
3.4

Widgets Screen

From the Control Screen, touch the wrench button in the lower right corner of the screen.
This will take you to the first Widgets Screen (see Figure 9).
3.5

Setting the Sequence

Navigate to the first Widgets Screen. Touch the Sequences widget to take you to the
Sequences Screen (see Figure 29). Using the buttons on the screen, create the
sequences desired. Changes are saved by touching the save button in the lower right
hand corner. To exit without saving changes, touch the escape button in the lower left
hand corner.
3.6

Placing Samples in "Sample/Bypass"

Navigate to the first Widgets Screen. Touch the Bypass widget to access the Bypass
Line Screen (see Figure 12). Touch the indicator light next to the desired sample line
you wish to switch. The state of the sample line will update once indicator has been
touched. The Sample state indicates the sample point will be sampled. The Bypass state
indicates the sample point will be bypassed the next time its position is up in the
sequence. Changes are saved by touching the save button in the lower right hand
corner. To exit without saving changes, touch the escape button in the lower left hand
corner.
3.7

Setting Sample Timers

Navigate to the first Widgets Screen. Touch the Flush Times or Sample Times widget to
access the appropriate screen (see Figures 10 and 11). Touching the time on the
SPD 5.5.15
Sentry Equipment Corp | Sample Sequencer
5
®
.