II. Setting up the instrument
Avoid placing the instrument in locations exposed to direct sunlight, dust,
vibration, high temperature and high humidity.
Operating environment
Indoor use
Altitude: Max 2000 meters
Ambient temperature: 15°C to 35°C
Maximum relative humidity: 75% for temperature up to 31°C decreasing
linearly to 50% relative humidity at 40°C
Supply voltage fluctuations: Not to exceed ±10% of the normal voltage.
Pollution degree: 2 (in according with IEC60664)
Installation / Overvoltage category: 2 (in according with IEC60664)
Air pressure of 75kPa to 106 kPa
Avoid frost, dew, percolating water, and rain
III. Assembling the microscope

1. Verifying input voltage

The 5V microscope input can works with a broad range of USB power
sources. However, always use the provided automatic voltage selection
power adapter or own adapter, portable power device that are rated for
the voltage used in your area, microscope input and that has been
approved to meet local safety standards.
In order to prevent electric shock, always turn the power switch on the
power supply off before connecting the power cord.
Electrical specifications:
AC adapter: 100-240V 50/60 Hz (Included)
Microscope Input: Type-C 5V/2A by AC adapter or portable power device
(not included)

2. Illumination

LED illumination
LED Illumination Specification:
6