June 2022
1.3
Before removing the 3800 Series Electronics
1.3.1
Procedure before removing the 3800 Series Electronics
Procedure
1. Before removing power from the meter being software to connect to
the meter.
2. Use the
to read the configuration from the meter and save it to a file on your
computer. An example for a file name might be "Meter Name, 3800
Series Final Config, 3-4-2013 10-15-02 AM.cfg". MeterLink will be
used later in this procedure (see
Electronics) to convert the configuration and download it to the
3810 Series electronics.
Loss of set-up data
The current meter configuration file must be saved before beginning
the 3810 Series electronics .
Failure to save the current configuration file can result in the loss of
important set-up data.
3 through 8 provide more detailed instructions on obtaining the
required information with respect to the communications set-up of
the 3800 Series electronics, however, if these settings are already
known, these steps can be skipped. Step 8 is for Frequency and
Diagnostic outputs. For reference, the known values should be
entered in the
Appendix
Series electronics. For any setting which is not known, the applicable
step(s) for obtaining that information can be used.
3. Use the
<meter name> you want to .
Upgrade kit instructions
Edit/Compare Configuration
CAUTION
Summary table of communications/output settings
B. This information will be required in the setup of the 3810
File|Meter Directory
a) Click the Direct button on the lower left corner of the Meter
Directory dialog box.
b) The Direct Connection Properties For dialog appears.
c) Obtain the Modbus ID (same as Comms Address).
Upgrade kit instructions
screen in MeterLink
Configure the 3810 Series
menu path, then select the
in
5